CD pH Meter incorporates the latest technological advancements. It has a
very high stability, reproducibility and a continuous duty cycle.
This laboratory pH Meter incorporates an advanced C-MOS FET input
operational amplifier and modular construction. It has a relative accuracy
of ±0.01 pH unit or ±1 mV based on a low point calibration,
repeatability of ±0.01 pH unit or ±1 mV and a stability (drift/24
hrs) of less than 0.01 pH.
It provides a digital LED (Model APX 175 E) and LCD (Model APX
175 C) display of pH 0.00 to pH 14.00 in 0.01 pH unit or milli Volts from 0 to
±1999 to nearest 1 mV.
Automatically indicated on the digital display are decimal point, pH/mV mode
and mV polarity.
The following controls are provided on the front panel:
Push button controls
:
- Standby/Read position
- pH / mV measurements
- Manual / Automatic Temperature Compensation.
|
Dial Controls:
- Temperature compensation control to read from -10°C to 130°C.
- Slope control for correcting Electrode slope
- Asymmetry control - adjustable span ± 1 pH by a 10 turn potentiometer,
in steps of 0.01 pH.
|
The CD pH Meter is a light-weight instrument with aluminium sheet metal housing
and is epoxy painted.
A unique feature is the two-dimentional electrode stand with a flexible
spring balanced arm. A third socket is also provided for the optional auto
temperature compensator.
Along with pH Meter we provide a combined electrode, electrode stand,
electrode holder, buffers, wash bottle, dust protective cover and handy
Instruction manual.
The ON / OFF Mains switch is mounted on the back panel with a status flag. Also
mounted on the back panel are plug-in sockets for Electrodes, Automatic
Temperature Compensator, Recorder Output
|
SPECIFICATIONS
|
MODEL APX 175 E
|
MODEL APX 175 C
|
|
Display Range
|
3½ Digit LED
7 Segment 0.5" Height
0.00 to 14 pH / 0 to ± 1999 mV
|
3½ Digit LCD
7 Segment 0.5" Height
0.00 to 14 pH / 0 to ± 1999 mV
|
|
Relative Accuracy
|
±0.01 pH / ±1.0 mV
|
±0.01 pH / ±1.0 mV
|
|
Repeatability
|
±0.01 pH / ±1.0 mV
|
±0.01 pH / ±1.0 mV
|
|
Stability (dirft/24 hrs)
|
<0.01 pH after warm up period of 30 minutes
non-cumulative
|
<0.01 pH after warm up period of 30 minutes
non-cumulative
|
|
Slope Adjustment
|
|
|
|
Asymmetry Potential
|
Adjustable ±1 pH by a 10 turn potentiometer
|
Adjustable ±1 pH by a 10 turn potentiometer
|
Temperature Compensation
Automatic(with Compensator probe APX 160)
|
-10°C to 130°C
|
-10°C to 130°C
|
|
Manual
|
-10°C to 130°C
|
-10°C to 130°C
|
|
Input Impedance
|
>10
14
Ohms
|
>10
14
Ohms
|
|
Polarizing current
|
10 µAmps
|
10 µAmps
|
|
Recorder output
|
0 to 100 mV full scale
|
0 to 100 mV full scale
|
|
Battery operation
|
No
|
Yes, 6 nos. 1.5V Cells Type Eveready 1050
|
|
Auto switch over from mains to battery during power
failure
|
No
|
Yes
|
|
Mains Supply
|
230V AC 50Hz ±10%
|
230V AC 50Hz ±10%
|