| Analog
Input |
|
|
Configuration : |
Unipolar differential
|
|
Bias current : |
<10
- 12
amps for both pH and reference electrode Inputs
|
|
Input Impedance : |
>10
12
Ohms for both pH and reference electrode Inputs
|
| Analog
Output |
|
|
Full Scale Measuring Ranges : |
(pH 0 - pH 10) / (pH 0 - pH 5)
(pH 1 - pH 11) / (pH 1 - pH 6)
(pH 2 - pH 12) / (pH 2 - pH 7)
(pH 4 - pH 14) / (pH 4 - pH 9)
(pH 7 - pH 14) / (pH 7 - pH 12)
Corresponding ranges in millivolts. 20 ranges possible between -1000 to +1000 mV.
|
|
Output Current : |
0 - 20 mA DC Dead Zero ; convertable to 0 - 5 V DC
4 - 20 mA DC Live Zero ; convertable to 1 - 5 V DC
|
|
Max. Load resistance : |
500 Ohms (10 V) at 300 meters.
|
| Signal Conditioning |
|
|
Temperature Compensation : |
Automatic between -10°C to +150°C using PT100 Platinum Thermocompensator "APX 140"
|
|
Asymmetry correction : |
± 60mV / ± 120 mV ± 1 pH / ± 2 pH
|
|
Slope Correction : |
90 to 105% at 25°C of Nernst value
|
| Accuracy & Stability |
|
|
Linearity : |
Linearity error less than ± 1% FSD
|
|
Reproducibility : |
Better than ± 0.1% FSD
|
|
Stability : |
Drift after 1 hour of operation better than 0.01% FSD in 24 hours With fluctuating ambient temperature better than ± 0.25% FSD/10°C
|
|
Common mode range : |
Max. ± 10V DC ± 10V AC (Peak value)
|
|
Influence of load variation : |
< 0.1% FSD for 500 Ohms change.
|
|
Power : |
230 Volts AC single phase / 110 volts AC single phase 24V DC +20% -10% 3 Watts maximum
|
| Environmental, Physical & Reliability |
|
|
Temperature Range : |
Operating 0°C to 60°C Storage -25°C to 85°C
|
|
Relative Humidity : |
0 to 95% non-condensing
|
|
Warm up Time : |
5 minutes
|
|
Over voltage : |
Power : 260 volts max. 200 volts min. Input : 2 volts max.
|
|
Case : |
Die Cast aluminium, weather and splash proof
|
|
Mounting : |
Surface mounting or Panel mounting on three bolts.
|
Digital Display |
|
|
Local Digital Readout : |
3 ½ digit (LED or LCD) 0.5" height 7 segment display
|