CD Digital Conductivity Meters has features - high stability, reproducibility,
and continuous duty cycle.
The model APX 185 LABORATORY conductivity meter incorporates a J FET
controlled oscillator and EPROM switching, with automatic ability to choose
100Hz to 1000Hz operation. It has a relative accuracy of ±1 least digit of
selected range, repeatability of ±1 least digit and stability
(drift/24hrs) after warm up of 30 minutes of less than 1 least digit.
SALIENT FEATURES
- WIDE MEASUREMENT RANGE
From 200nS/cm to 200mS/cm (1S = 1 mho)
- HIGH SENSITIVITY
100pS/cm using 0.01 cell constant.
- DIRECT DISPLAY OF CONDUCTANCE
The display is in conductance mS/cm and µS/cm (conductivity x cell
constant) eliminating cell constant determination and user multiplication.
- AUTOMATIC TEMPERATURE COMPENSATION
With an optional NTC probe, the conductivity variations of the sample with
temperature compensated.
|
- RATIOMETRIC MEASUREMENT
Using two conductivity cells the instruments can be used as a conductivity
comparator for quality control "GO-NOGO" tests.
- AUTOMATIC SELECTION OF OPTIMUM FREQUENCY
The instrument automatically chooses the internal frequency of 100Hz or
1000Hz to suit the measurement range for very high accuracy readings.
- DIGITAL READOUT
Bright 3 ½ digit 7 segment LED display for ease of observation for
readings from 0.0001µS/cm to 19.99 mS/cm
|
UNIQUE FEATURES
Measurements in conductivity are done for determining the conductance of
samples (ie. conductivity displayed on the usual meters multiplied by the cell
constant of cell). The user has to determine the cell constant accurately from
time to time and use this to multiply the displayed reading and obtain
CONDUCTANCE in S/cm. the CD CONDUCTIVITY METER accurately
determines the cell constant during calibration, multiplies and displays
CONDUCTANCE DIRECTLY ensuring the highest accuracy possible.
Another unique feature is the Two dimensional electrode stand with a
flexible spring balanced arm.
Along with the Conductivity Meter we provide a 1.0 cell K sensor, electrode
stand, electrode holder, wash bottle, calibrating standards, a dust protective
cover and a handy Plastic Instruction manual.
The CD Conductivity meter is a light-weight instrument with aluminium sheet
metal housing and is epoxy painted.
|
SPECIFICATIONS
|
MODEL APX 185E
|
|
Display Range
|
3½ Digit LED
7 Segment 0.5" Height
0 - 20µS; 0 - 200µS
0 - 2mS; 0 - 20mS
and RATIOMETRIC 0.000 to 1.000
|
|
Relative Accuracy
|
± 1 Last digit of Range displayed
|
|
Minimum / Maximum Reading
|
0.0001 µS/cm with 0.01 Cell K
199.9 mS/cm with 10.0 Cell K
|
|
Repeatability
|
± 1 Last digit of Range displayed
|
|
Linearity
|
Better than 1% of Range displayed
|
|
Stability (drift/24 hrs)
|
<1 Digit after warm up period of 30 minutes
non-cumulative
|
|
Temperature Compensation
|
Manual 0 °C to 100 °C on Front Panel
Automatic with NTC Temperature Compensator probe or with Reference Cell
|
|
Recorder Output
|
Analog 0-100mV Full Scale
|
|
Mains Supply
|
230V 50Hz AC ±10%
|
|
Environmental Conditions
|
Operating Temperature -20°C to 55°C, Humidity 90°C
RH at 40°C
|