 Digital Conductivity Meter : Model APX 185E
 

CD Digital Conductivity Meters has features - high stability, reproducibility, and continuous duty cycle.

The model APX 185 LABORATORY conductivity meter incorporates a J FET controlled oscillator and EPROM switching, with automatic ability to choose 100Hz to 1000Hz operation. It has a relative accuracy of ±1 least digit of selected range, repeatability of ±1 least digit and stability (drift/24hrs) after warm up of 30 minutes of less than 1 least digit.

SALIENT FEATURES

  • WIDE MEASUREMENT RANGE
    From 200nS/cm to 200mS/cm (1S = 1 mho)

  • HIGH SENSITIVITY
    100pS/cm using 0.01 cell constant.

  • DIRECT DISPLAY OF CONDUCTANCE
    The display is in conductance mS/cm and µS/cm (conductivity x cell constant) eliminating cell constant determination and user multiplication.

  • AUTOMATIC TEMPERATURE COMPENSATION
    With an optional NTC probe, the conductivity variations of the sample with temperature compensated.

  • RATIOMETRIC MEASUREMENT
    Using two conductivity cells the instruments can be used as a conductivity comparator for quality control "GO-NOGO" tests.

  • AUTOMATIC SELECTION OF OPTIMUM FREQUENCY
    The instrument automatically chooses the internal frequency of 100Hz or 1000Hz to suit the measurement range for very high accuracy readings.

  • DIGITAL READOUT
    Bright 3 ½ digit 7 segment LED display for ease of observation for readings from 0.0001µS/cm to 19.99 mS/cm
UNIQUE FEATURES

Measurements in conductivity are done for determining the conductance of samples (ie. conductivity displayed on the usual meters multiplied by the cell constant of cell). The user has to determine the cell constant accurately from time to time and use this to multiply the displayed reading and obtain CONDUCTANCE in S/cm. the CD CONDUCTIVITY METER accurately determines the cell constant during calibration, multiplies and displays CONDUCTANCE DIRECTLY ensuring the highest accuracy possible.

Another unique feature is the Two dimensional electrode stand with a flexible spring balanced arm.

Along with the Conductivity Meter we provide a 1.0 cell K sensor, electrode stand, electrode holder, wash bottle, calibrating standards, a dust protective cover and a handy Plastic Instruction manual.

The CD Conductivity meter is a light-weight instrument with aluminium sheet metal housing and is epoxy painted.

SPECIFICATIONS MODEL APX 185E
Display Range 3½ Digit LED
7 Segment 0.5" Height
0 - 20µS;  0 - 200µS
0 - 2mS;  0 - 20mS
and RATIOMETRIC 0.000 to 1.000
Relative Accuracy ± 1 Last digit of Range displayed
Minimum / Maximum Reading 0.0001 µS/cm with 0.01 Cell K
199.9 mS/cm with 10.0 Cell K
Repeatability ± 1 Last digit of Range displayed
Linearity Better than 1% of Range displayed
Stability (drift/24 hrs) <1 Digit after warm up period of 30 minutes non-cumulative
Temperature Compensation Manual 0 °C to 100 °C on Front Panel
Automatic with NTC Temperature Compensator probe or with Reference Cell
Recorder Output Analog 0-100mV Full Scale
Mains Supply 230V 50Hz AC ±10%
Environmental Conditions Operating Temperature -20°C to 55°C, Humidity 90°C RH at 40°C


| home | products | about us | contact us |