 CLINICAL DOSIMETER
 

The measurement of Patient Dose during Brachytherapy is a standard practice followed by hospitals all over the world.

Clinidose enables reliable Rectal Dose measurements with high degree of safety to the Patients.

Clinidose is a simple instrument to operate. Adjusting the digital display to zero ( in the absence of Radiation) is the only control that the instrument needs.

Clinidose has a built-in rechargeable Ni-Cd Battery supply. For protection from shocks to the patients, all dose measurements are performed while operating the unit on Batteries.

SPECIFICATIONS Detector Probe with Ionization Chamber Electrometer
Radiation Detected X –ray 80 to 250 kV
Gamma Ray with build–up material
Chamber Volume 0.2 CC
Chamber Wall Material Tufnol  (Air Equivalent)
Chamber Location Inside a graduated, Chrome plated Brass Probe
Probe Diameter 7.5 mm
Probe length 310mm
Connectors BNC connectors Signal and Bias voltages
Input Impedance >10 14 Ohms >10 14 Ohms
Cable Length 2.5 Meter (extra length optional)
Bias Voltage 15 Volts 15 Volts
Stability (Leakage) Less than 5 x 10-14A Less than 5 x 10-14A
Accuracy Better than +/- 10 %
Reproducibility +/- 3 %
Modes Dose Rate
Digital Display 3½ Digit LCD.  0000 to 1999 MU  (Calibrated in cGy / hr )
Recorder output 0 to 100 mV full scale
Power Ni-Cd Battery pack
Battery Charger Inbuilt 115V/230V AC 60Hz/50Hz ±10%
Carry Case Yes Yes