CD Hightech Pvt. Ltd.

Home Page
Products
About us
Contact us



 Brachytherapy Calibrator
 

Use of Radio-nuclides with in-house verification of source strength is advisable due to the accuracy required for maximizing tumor cell death and minimizing healthy tissue damage. Well-type ionization chambers make it easy to establish source calibration procedures. For accreditation of institutes, some countries have made usage of Well Chamber a mandatory requirement.



The Brachytherapy Calibrator has been developed by CD Hightech in collaboration with Bhabha Atomic Research Centre, India and is tested at various leading hospitals. The unit can be used for the calibration of Isotopes in both HDR and LDR applications.The source holders provided would enable high reproducibility of the source position in the chamber. Tests conducted at hospitals confirm excellent performance of the unit.





SPECIFICATIONS Well chamber Electrometer
Range

0.1mCi to 20Ci

Active Volume 250 CC



Source Holders Ir-192 Wires, Coils,
LDR & MDR Applicators,
Seeds etc.


Connectors BNC connectors Signal and Bias voltages
Input Impedance >10 14 Ohms >10 14 Ohms
Cable Length 1 Meter (extra length optional)
Bias Voltage 160 and 320 Volts 160 and 320 Volts
Stability (Leakage) Less than 5 x 10-14A Less than 5 x 10-14A
Relative Accuracy Better than +/- 1%
Repeatability +/- 0.5%
Modes Charge & Current
Digital Display
3½ Digit LCD displaying
nano-Coulombs/ pico-Amperes
Recorder output 0 to 100 mV full scale
Mains Supply 115V/230V AC 60Hz/50Hz ±10%
Carry Case Yes Yes


| home | products | about us | contact us |

© CD Hightech Pvt. Ltd. 1999

Prev Product Product Index Next Product